RAPID: A routine assurance pipeline for imaging of diffusion
Journal article
Santis, Silvia De, Evans, C. J. and Jones, D. K.. (2013). RAPID: A routine assurance pipeline for imaging of diffusion. Magnetic Resonance in Medicine. 70(2), pp. 490 - 496. https://doi.org/10.1002/mrm.24465
Authors | Santis, Silvia De, Evans, C. J. and Jones, D. K. |
---|---|
Abstract | Despite the widespread adoption of diffusion MRI techniques, there is still no consensus on a comprehensive quality assurance routine specific for diffusion acquisitions. We propose here a routine assurance pipeline for imaging of diffusion. The routine simply comprises diffusion-weighted acquisitions on a phantom; each repetition lasts less than 5 min and can be performed using a variety of isotropic test liquids. The proposed QA script checks for the linearity of G, the uniformity of Gmax across the field-of-view, the mutual agreement of gradient power across the three logical axes and the temporal stability. Optionally, the routine can correct for the mutual agreement of gradient power along the three axes, returning a set of gradient orientations to be used in data analysis. The effectiveness of the scheme in the presence of mismatched gradient amplitudes is reported using both simulations and in vivo data. The script is freely available online. |
Keywords | diffusion; quality assurance; gradient calibration; cross-terms |
Year | 2013 |
Journal | Magnetic Resonance in Medicine |
Journal citation | 70 (2), pp. 490 - 496 |
Publisher | John Wiley & Sons, Inc. |
ISSN | 0740-3194 |
Digital Object Identifier (DOI) | https://doi.org/10.1002/mrm.24465 |
Scopus EID | 2-s2.0-84881155270 |
Page range | 490 - 496 |
Publisher's version | File Access Level Controlled |
Place of publication | United States |
https://acuresearchbank.acu.edu.au/item/861v8/rapid-a-routine-assurance-pipeline-for-imaging-of-diffusion
Restricted files
Publisher's version
104
total views0
total downloads3
views this month0
downloads this month