A 12-Month Clinical Audit Comparing Point-of-Care Lactate Measurements Tested by Paramedics with In-Hospital Serum Lactate Measurements

Journal article


Kristi L. Swan, Toby Keene and Bronwyn J. Avard. (2018). A 12-Month Clinical Audit Comparing Point-of-Care Lactate Measurements Tested by Paramedics with In-Hospital Serum Lactate Measurements. Prehospital and Disaster Medicine. 33(1), pp. 36-42. https://doi.org/10.1017/S1049023X17007130
AuthorsKristi L. Swan, Toby Keene and Bronwyn J. Avard
Abstract

Objective
Prehospital point-of-care lactate (pLA) measurement may be a useful tool to assist paramedics with diagnosing a range of conditions, but only if it can be shown to be a reliable surrogate for serum lactate (sLA) measurement. The aim of this study was to determine whether pLA is a reliable predictor of sLA.

Methods
This was a retrospective study of adult patients over a 12-month period who had pLA measured by paramedics in an urban Australian setting and were transported by ambulance to a tertiary hospital where sLA was measured. Patients were excluded if they suffered a cardiopulmonary arrest at any time, had missing data, or if sLA was not measured within 24 hours of arrival. Levels of agreement were determined using methods proposed by Bland and Altman.

Results
A total of 290 patients were transported with a pLA recorded. After exclusions, there were 155 patients (55.0% male; age 71 [SD=18] years) remaining who had sLA recorded within 24 hours. Elevated pLA (>2.0mMol/L) was associated with sLA measurement (76.1% vs 23.9%; OR 3.18; 95% CI, 1.88-5.37; P<.0001). Median time between measurements was 89 minutes (IQR=75). Overall, median pLA was higher than sLA (3.0 [IQR=2.0] mMol/L vs 1.7 [IQR=1.3]; P<.001). Bland-Altman analysis on all participants showed a mean difference of 1.48 mMol/L (95% CI, -3.34 to 6.31). Normal pLA was found to be a true negative in 82.9% of cases, and elevated pLA was a true positive in 48.3% of cases. When the time between measurements was less than 60 minutes (n=25), normal pLA predicted normal sLA with 100% accuracy, with a false-positive rate of 18.2%. As time between measurements increased, accuracy diminished and the false-positive rate increased.

Conclusions
Overall, the level of agreement between pLA and sLA was poor. Accuracy of pLA diminished markedly as the time between the two measurements increased. It may be possible to use pLA as a screening tool; when considered this way, pLA performed much better, though larger prospective trials would be needed to confirm this.

Year2018
JournalPrehospital and Disaster Medicine
Journal citation33 (1), pp. 36-42
PublisherCambridge University Press
ISSN1049-023X
Digital Object Identifier (DOI)https://doi.org/10.1017/S1049023X17007130
Scopus EID2-s2.0-85041354448
Publisher's version
File Access Level
Controlled
Publication process dates
Deposited07 May 2021
Permalink -

https://acuresearchbank.acu.edu.au/item/8vzqw/a-12-month-clinical-audit-comparing-point-of-care-lactate-measurements-tested-by-paramedics-with-in-hospital-serum-lactate-measurements

Restricted files

Publisher's version

  • 42
    total views
  • 0
    total downloads
  • 3
    views this month
  • 0
    downloads this month
These values are for the period from 19th October 2020, when this repository was created.

Export as

Related outputs

Impacts on in-event, ambulance and emergency department services from patients presenting from a mass gathering event: A retrospective analysis
Ranse, Jamie, Lenson, Shane, Keene, Toby, Luther, Matt, Burke, Brandon, Hutton, Alison, Johnston, Amy NB and Crilly, Julia. (2019). Impacts on in-event, ambulance and emergency department services from patients presenting from a mass gathering event: A retrospective analysis. Emergency Medicine Australasia. 31(3), pp. 423 - 428. https://doi.org/10.1111/1742-6723.13194
The relationship between elevated prehospital point-of-care lactate measurements, intensive care unit admission, and mortality: A retrospective review of adult patients
Kristi L. Swan, Bronwyn J. Avard and Toby Keene. (2018). The relationship between elevated prehospital point-of-care lactate measurements, intensive care unit admission, and mortality: A retrospective review of adult patients. Australian Critical Care. 32, pp. 100-105. https://doi.org/10.1016/j.aucc.2018.02.006
Health service impact from mass gatherings: A systematic literature review
Ranse, Jamie, Hutton, Alison, Keene, Toby, Lenson, Shane, Luther, Matt, Bost, Nerolie, Johnston, Amy N. B., Crilly, Julia, Cannon, Matt, Jones, Nicole, Hayes, Courtney and Burke, Brandon. (2017). Health service impact from mass gatherings: A systematic literature review. Prehospital and Disaster Medicine. 32(1), pp. 71 - 77. https://doi.org/10.1017/S1049023X16001199
Characteristics and outcomes of patients assessed by paramedics and not transported to hospital: A pilot study
Keene, Toby, Davis, Megan and Brook, Carol. (2015). Characteristics and outcomes of patients assessed by paramedics and not transported to hospital: A pilot study. Australasian Journal of Paramedicine. 12(2), pp. 1 - 8.
Immune and inflammatory responses of Australian firefighters after repeated exposures to the heat
Walker, Anthony, Keene, Toby, Argus, Christos, Driller, Matthew, Guy, Joshua H. and Rattay, Ben. (2015). Immune and inflammatory responses of Australian firefighters after repeated exposures to the heat. Ergonomics. 58(12), pp. 2032 - 2039. https://doi.org/10.1080/00140139.2015.1051596