Czerniec, Sharon Anne, Ward, Leigh, Lee, Mi-Joung, Refshauge, Kathryn, Beith, Jane and Kilbreath, Sharon. (2011). Segmental measurement of breast cancer-related arm lymphoedema using perometry and bioimpedance spectroscopy. Supportive Care in Cancer. 19(5), pp. 703 - 710. https://doi.org/10.1007/s00520-010-0896-8
AuthorsCzerniec, Sharon Anne, Ward, Leigh, Lee, Mi-Joung, Refshauge, Kathryn, Beith, Jane and Kilbreath, Sharon
Abstract

Purpose: To determine if bioimpedance spectroscopy (BIS) could detect localised lymphoedema of the arm and to compare BIS measurements with equivalent measures of limb volume by perometry.

Methods: Women with mild to severe upper limb lymphoedema (n = 29) and women with no history of lymphoedema (n = 11) participated. Commencing at the ulnar styloid of the wrist, 4 × 10 cm segment measurements were made of each arm using both BIS and perometry.

Results: Average BIS inter-limb ratios for the total arm and each arm segment were higher than comparable perometry measures in women with lymphoedema, but similar to perometry measures for women without lymphoedema. Limits of agreement analysis showed that the mean difference between methods varied according to segment measured, ranging from 8.5% for the uppermost segment of the arm to 16.6% for the forearm segment just below the elbow. For all limb segments, there was a positive bias towards BIS measurements, which increased as lymphoedema severity increased.

Conclusion: BIS can be used for localised measurement of lymphoedema. Because it is specific to extracellular fluid, BIS is more sensitive to localised lymphoedema than perometry.

Year2011
JournalSupportive Care in Cancer
Journal citation19 (5), pp. 703 - 710
ISSN0941-4355
Digital Object Identifier (DOI)https://doi.org/10.1007/s00520-010-0896-8
Scopus EID2-s2.0-79958199593
Page range703 - 710
Research GroupSchool of Allied Health
Permalink -

https://acuresearchbank.acu.edu.au/item/8678v/segmental-measurement-of-breast-cancer-related-arm-lymphoedema-using-perometry-and-bioimpedance-spectroscopy

  • 102
    total views
  • 0
    total downloads
  • 2
    views this month
  • 0
    downloads this month
These values are for the period from 19th October 2020, when this repository was created.

Export as

Related outputs

Estimation of Arm Adipose Tissue Quotient Using Segmental Bioimpedance Spectroscopy
Ruben Buendia, T. Essex, Sharon Kilbreath, Sharon Czerniec, Elizabeth Dylke and Leigh C. Ward. (2018). Estimation of Arm Adipose Tissue Quotient Using Segmental Bioimpedance Spectroscopy. Lymphatic Research and Biology. 16(4), pp. 377-384. https://doi.org/10.1089/lrb.2017.0050
Does lymphedema severity affect quality of life? Simple question. Challenging answers
Teresa S. Lee, Carol M. Morris, Sharon Czerniec and Andrea J. Mangion. (2018). Does lymphedema severity affect quality of life? Simple question. Challenging answers. Lymphatic Research and Biology. 16(1), pp. 85-91. https://doi.org/10.1089/lrb.2016.0049
A patient-based self-examination survey for staging the severity of lymphoedema
Morris, Carol M., Lee, Teresa S., Czerniec, Sharon A. and Mangion, Andrea J.. (2017). A patient-based self-examination survey for staging the severity of lymphoedema. Journal of Lymphoedema. 12(1), pp. 10 - 15.
Breast Cancer-Related Arm Lymphedema: Fluctuation over Six Months and the Effect of the Weather
Czerniec, Sharon Anne, Ward, Leigh and Kilbreath, Sharon. (2016). Breast Cancer-Related Arm Lymphedema: Fluctuation over Six Months and the Effect of the Weather. Lymphatic Research and Biology. 14(3), pp. 148 - 155. https://doi.org/10.1089/lrb.2015.0030
Assessment of segmental arm soft tissue composition in breast cancer-related lymphedema : A pilot study using dual energy X-ray absorptiometry and bioimpedance spectroscopy
Czerniec, Sharon Anne, Ward, Leigh, Meerkin, Jarrod and Kilbreath, Sharon. (2015). Assessment of segmental arm soft tissue composition in breast cancer-related lymphedema : A pilot study using dual energy X-ray absorptiometry and bioimpedance spectroscopy. Lymphatic Research and Biology. 13(1), pp. 33 - 39. https://doi.org/10.1089/lrb.2014.0033
Assessment of breast cancer-related lymphedema : A comparison of moisture meter and spot bioimpedance measurement
Czerniec, Sharon Anne, Ward, Leigh and Kilbreath, Sharon. (2015). Assessment of breast cancer-related lymphedema : A comparison of moisture meter and spot bioimpedance measurement. Lymphatic Research and Biology. 13(1), pp. 10 - 19. https://doi.org/10.1089/lrb.2014.0032
Estimation of limb adiposity by bioimpedance spectroscopy in lymphoedema
Ward, Leigh, Essex, T., Gaw, R., Czerniec, Sharon, Dylke, Elizabeth, Abell, B. and Kilbreath, Sharon. (2013). Estimation of limb adiposity by bioimpedance spectroscopy in lymphoedema. Journal of Physics: Conference Series. 434(1), pp. 1 - 4. https://doi.org/10.1088/1742-6596/434/1/012062
Assessment of bilateral limb lymphedema by bioelectrical impedance spectroscopy
Ward, Leigh, Winall, Ann, Isenring, Elizabeth, Hills, Andrew, Czerniec, Sharon, Dylke, Elizabeth and Kilbreath, Sharon. (2011). Assessment of bilateral limb lymphedema by bioelectrical impedance spectroscopy. International Journal of Gynecological Cancer. 21(2), pp. 409 - 418. https://doi.org/10.1097/IGC.0b013e31820866e1
Reference ranges for assessment of unilateral lymphedema in legs by bioelectrical impedance spectroscopy
Ward, L., Dylke, E., Phty, M., Czerniec, S., Isenring, E. and Kilbreath, S.. (2011). Reference ranges for assessment of unilateral lymphedema in legs by bioelectrical impedance spectroscopy. Lymphatic Research and Biology. 9(1), pp. 43 - 46. https://doi.org/10.1089/lrb.2010.0024
Confirmation of the reference impedance ratios used for assessment of breast cancer-related lymphedema by bioelectrical impedance spectroscopy
Ward, Leigh, Dylke, Elizabeth, Czerniec, Sharon Anne, Isenring, Elisabeth and Kilbreath, Sharon. (2011). Confirmation of the reference impedance ratios used for assessment of breast cancer-related lymphedema by bioelectrical impedance spectroscopy. Lymphatic Research and Biology. 9(1), pp. 47 - 51. https://doi.org/10.1089/lrb.2010.0014
Reliability and concurrent validity of the perometer for measuring hand volume in women with and without lymphedema
Lee, Mi-Joung, Boland, Robert, Czerniec, Sharon Anne and Kilbreath, Sharon. (2011). Reliability and concurrent validity of the perometer for measuring hand volume in women with and without lymphedema. Lymphatic Research and Biology. 9(1), pp. 13 - 18. https://doi.org/10.1089/lrb.2010.0021
Reference ranges for assessment of unilateral lymphedema in legs by bioelectrical impedance spectroscopy
Ward, Leigh, Dylke, Elizabeth, Czerniec, Sharon Anne, Isenring, Elisabeth and Kilbreath, Sharon. (2011). Reference ranges for assessment of unilateral lymphedema in legs by bioelectrical impedance spectroscopy. Lymphatic Research and Biology. 9(1), pp. 43 - 46. https://doi.org/10.1089/lrb.2010.0024
Confirmation of the reference impedance ratios used for assessment of breast cancer-related lymphedema by bioelectrical impedance spectroscopy
Ward, Leigh, Dylke, Elizabeth, Czerniec, Sharon, Isenring, Elisabeth and Kilbreath, Sharon. (2011). Confirmation of the reference impedance ratios used for assessment of breast cancer-related lymphedema by bioelectrical impedance spectroscopy. Lymphatic Research and Biology. 9(1), pp. 47 - 51. https://doi.org/10.1089/lrb.2010.0014
Segmental measurement of breast cancer-related arm lymphoedema using perometry and bioimpedance spectroscopy
Czerniec, Sharon, Ward, Leigh, Lee, Mi-Joung, Refshauge, Kathryn, Beith, Jane and Kilbreath, Sharon. (2011). Segmental measurement of breast cancer-related arm lymphoedema using perometry and bioimpedance spectroscopy. Supportive Care in Cancer. 19(5), pp. 703 - 710. https://doi.org/10.1007/s00520-010-0896-8
Reliability and concurrent validity of the perometer for measuring hand volume in women with and without lymphedema
Lee, Mi-Joung, Boland, Robert, Czerniec, Sharon and Kilbreath, Sharon. (2011). Reliability and concurrent validity of the perometer for measuring hand volume in women with and without lymphedema. Lymphatic Research and Biology. 9(1), pp. 13 - 18. https://doi.org/10.1089/lrb.2010.0021
Assessment of bilateral limb lymphedema by bioelectrical impedance spectroscopy
Ward, Leigh, Winall, Ann, Isenring, Elisabeth, Hills, Andrew, Czerniec, Sharon Anne, Dylke, Elizabeth and Kilbreath, Sharon. (2011). Assessment of bilateral limb lymphedema by bioelectrical impedance spectroscopy. International Journal of Gynecological Cancer. 21(2), pp. 409 - 418. https://doi.org/10.1097/IGC.0b013e31820866e1